![Honeywell SmartLine SLN 700 82 Series User Manual Download Page 47](http://html.mh-extra.com/html/honeywell/smartline-sln-700-82-series/smartline-sln-700-82-series_user-manual_154963047.webp)
Revision 1
SLN 700 SmartLine NCR Level Transmitter User’s Manual
Page 39
Error Codes:
Following are the details of error codes with respective resolution. The error codes will show
up on display screen during malfunction/error condition in device
Error
Code
Description
Solutions
1
Communication error between display
and module
Need new display module
2
EEPROM Error
Return to Factory
3
PLL Failed, RF circuit failure
Return to Factory
4
Internal Operation Error or use during
development, internal use only)
Return to Factory
5
Internal Command Error for use during
development, internal use only)
Return to Factory
6
Authentication Failed
Return to Factory
7
SysBusy
Restart, if problems persist return unit to
Factory
8
Low Power
Low power supply. Check power supply
9
Dsp Error
Return to Factory
10
Internal Command Parse Error (for use
during development, internal use only)
Return to Factory
11
AD Error
AD sampling error, Return to Factory
12
Calibration Flash Error
RF Calibration Flash Error, Return to Factory
13
Dsp Authentication
Return to Factory
14
No Signal Error
No signal back, will see no reflective echo on
echo curve
Check antenna aiming
Check if the application is having excessive
vapor/steam