![Honeywell SmartLine SLN 700 82 Series User Manual Download Page 39](http://html.mh-extra.com/html/honeywell/smartline-sln-700-82-series/smartline-sln-700-82-series_user-manual_154963039.webp)
Revision 1
SLN 700 SmartLine NCR Level Transmitter User’s Manual
Page 31
5 System
The system menu contains basic information about the meter production,
such as
Info, Software Version, Latching Mode, Write Protect, etc.
When the LCD displays the main menu, press
key to move the
arrow to the information item. Display in
LCD is as follows.
Press
to enter the information display function. Display in LCD
is as follows.
5.1 Info
Product Type
Product Serial Number (ID)
Production date (PD)
5.2 Software
Version
5.3 Latching
Mode
Press
key. Display in LCD is as follows.
Latching Mode: This parameter allows selection of critical error
behavior.
Latching: The transmitter will remain in a critical error state until a user
performs a hardware / software reset.
Non-Latching: The transmitter exits critical error state automatically
when causes of the critical error have been resolved.
5.4 Write
Protect
If switching Yes to No, go to Unlock Password Input Menu
If switching No to Yes, go to set lock Password Input Menu.
Basic Settings
Display
Diagnostics
Advanced Settings
System
Information