Setup and Use
11
2.7 Verify Calibration
For best results, verify the calibration daily.
Note: When not using the original transducer, calibrate the Model NK before use
1. Obtain the following values from the transducer calibration data sheet:
•
Shunt-cal output (or “shunt-cal factor”) in mV/V
•
Calibration factor (or “full-scale output”) in mV/V
•
Full-scale engineering units (or “capacity”) in psig, lbs, etc.
2. Perform this calculation and record the value.
Note: Place a decimal point in the
full-scale engineering units
value in the equation in
the same place as set in the display.
3. With no load on the transducer, press and hold the
shunt cal
button.
Compare the displayed value with the result of step 2. If the values are
not equal, adjust the
fine span
potentiometer according to the procedure
below.