Calibration
20
2.
Fine span
—With the known load applied, use a small screwdriver to turn
the
fine span
potentiometer until the display reads exactly
19999
(or the
scaled-down value).
3.
Scaling
—This control scales the display to the desired units.
With the known load applied, use a small screwdriver to turn the
scaling
potentiometer until the display reads the desired full-scale engineering
units (found on the transducer’s calibration sheet or the result of step 1c).
After scaling, return the panel to the case and tighten the four captive fasteners. If
the display does not read
0000
, fine tune the unit again according to the procedure
on page 12.