Setup and Use
8
2.5 Turn On the Unit
Flip the
on/off
toggle switch on the front panel to
on
and allow the Model NK to
stabilize for a few seconds.
1. Make sure that no load is applied to the transducer, then press and hold
the
shunt cal
button.
2. If the number on the display drifts, the unit is not yet stable.
3. Release the
shunt cal
button and try again in a few seconds.
2.6 Set the Decimal Point Location
The Model NK is factory set to the correct decimal point location for its original
transducer. As a stand-alone product, it is factory set to xxx.xx. You can easily
change the decimal point location to match your load and desired display.
Each decimal point has two terminals that must be connected. For example, to set
the decimal point to xxxx.x for a 1000 psig load, jumper the first and second
terminals (from the front of the board).
For smaller loads (such as 200 psig), setting the decimal point to xxx.xx may
provide higher resolution, but may also result in an unstable last digit. For a stable