Calibration
23
4.
Scaling
—This control scales the display to the desired units. Determine
the scaled shunt-cal value using the following formula:
With no load on the transducer, press and hold the
shunt cal
button. Use a
small screwdriver to adjust the
scaling
potentiometer until the display
reads the scaled shunt-cal engineering units.
After scaling, return the panel to the case and tighten the four captive fasteners. If
the display does not read
0000
, fine tune the unit again according to the procedure
on page 12.