Calibration
14
3 Calibration
3.1 When to Calibrate
Calibrate the Model NK if you are not using the original transducer and each time
you change transducers.
3.2 Methods of Calibration
Two methods of calibration are available:
standard
and
shunt
. Use standard
calibration when you can place a known load on the transducer. Use shunt
calibration to simulate the load when you cannot apply a known load or for quick
field setup.