
Scale Calibration Procedure (lbs. or kg) with Remote Display
1. Temporarily Remove the platter and place it in a safe location.
Note:
It is the responsibility of the owner of the scale to confirm compliance with the relevant Weights
and Measures statutes and regulations applicable in your area by checking with the appropriate
government agency before placing a newly calibrated unit into service or removing any official
seals.
2. If this is a currently installed scanner/scale in need of calibration, cut and remove the calibration switch
cover seal. If this is a new installation, cut and remove the factory-applied adhesive seal.
Follow all Electro-Static Discharge (ESD) procedures when exposing internal scanner/scale
components.
3. Remove the M3 screw securing the calibration switch/button cover. Place the cover and screw in a safe
location. Verify that the scale calibration switch is in the Run position.
Figure 38. Calibration Switch/Button Cover Removal
46
Summary of Contents for ms2420
Page 1: ...MS2420 Bar Code Scanner Avery Scale Installation and User s Guide ...
Page 10: ...6 ...
Page 13: ...Base Model Characteristics Components Figure 7 MS2420 Components 9 ...
Page 32: ...28 ...
Page 62: ...58 ...
Page 68: ...64 ...
Page 76: ...72 ...
Page 78: ...74 ...
Page 80: ...76 ...
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ...Honeywell Scanning Mobility 90 Coles Road Blackwood NJ 08012 00 05312 Rev B September 2009 ...