IQ Force™ Dock Reference Manual
34
Test Sequence
The dock will now verify the PIN in any instrument that interfaces
with it.
10. For every other dock in the system, access the dock’s advanced
settings, enter the PIN, click on
Check PIN
and click the
Set
Config
button.
With PIN lock enabled, only instruments with a PIN matching the dock’s
PIN will be processed.
4. Dock Use
Once the IQ Force is connected to the gas cylinder(s) and has been properly
configured, it is ready to accept a IQ Force instrument.
The IQ Force must be in one of the following states before it is placed in
the dock:
1.
Instrument is on and displaying the gas monitoring screen.
2.
From the OFF state, press and hold the Mode button until
PC
Connect
appears on the screen.
NOTE
If the IQ Force has just been turned on, wait at least three
minutes after the conclusion of the startup sequence to
allow the readings to fully stabilize before placing the
instrument in the dock.
NOTE
If a sensor has just been replaced, it must be allowed to
stabilize before placing the instrument in the dock. See
the sensor replacement section of the IQ Force Reference
Manual for further information on sensor stabilization.
Once the dock recognizes the instrument, it will show the following screens.
Once the connection is made, the IQ Force will display Connected, followed
by the instrument type, software version, and recognized sensors.
If the dock recognizes the IQ Force but is unable to maintain the connection,
the following screen will be shown:
See
Section 4.2
for more information on connection issues
4.1 Test Sequence
Once the instrument has been recognized, the settings from the IQ Force’s
assigned IQ template will be applied. See the Database Manager Reference
manual for details on templates.
Summary of Contents for IQ Force
Page 1: ...IQ Force Dock Reference Manual ...
Page 2: ......
Page 54: ...IQ Force Dock Reference Manual 46 Help Menu ...
Page 57: ...49 Calibration Frequency Recommendation IQ Force Dock Reference Manual Appendix ...
Page 58: ......
Page 59: ......
Page 60: ...50109191 140 Revision 1 August 2012 2012 Honeywell Analytics All Rights reserved ...