INTELLIDOX DOCKING MODULE USER MANUAL || UPDATE FIRMWARE
HONEYWELL
PAGE 89 OF 119
Transferring Detector Firmware via Safety Suite Device COnfigurator
When a module or group of modules is connected to a network, you may transfer the
detector firmware update file to multiple selected modules via Safety Suite Device
Configurator software. For more information, refer to the Safety Suite Device Configurator
operator manual.
1. Download and save the firmware update file to a PC or network drive. Do not
rename the file.
2. Start Safety Suite Device Configurator software, and then log in as an
administrator. Select
Configure Devices via IntelliDoX
on the
Devices
menu.
The
IntelliDoX Configuration Selection
dialog box is displayed.
3. Select a configuration file and click
OK
. The
IntelliDoX Device Configuration
dialog box is displayed.
4. Click
Bootloader
. The
IntelliDoX Selection
dialog box is displayed.
5. Click
Choose File
. The
Choose Firmware File To Upload
dialog box is
displayed. Select the downloaded file, and then click
Open
. The
IntelliDoX
Selection
dialog box is displayed.
6. Select one or more IntelliDoX modules, and then click
OK
. A progress screen is
displayed while the firmware update file is automatically transferred to the selected
modules.
7. When the transfer is complete, click
OK
to close the dialog box.
Updating the Detector Firmware via Safety Suite Device Configurator
Modules must be configured via Safety Suite Device Configurator software to enable
firmware updates on insertion for compatible detectors. For more information, see
Configure Module and Detector Settings
.
1. Insert a compatible detector into the module.
Detector identification
is displayed
on the LCD.
2. If the module is configured to perform a firmware update and other procedures on
insertion and a firmware update is available, then the detector event logs are
automatically transferred to the module and the firmware update begins.
3. The LCD changes to yellow. A progress bar and the
Detector firmware upgrade
message are displayed.
4. After the upgrade is completed, the detector reboots. Alternating
Detector
rebooting
and
Do not open lid
messages are displayed.
5. When the reboot is successful, the LCD changes to green and the
Firmware
upgrade complete
message is displayed. Other procedures specified to occur
automatically on insertion are performed.
6. After all operations are completed, verify the firmware update.
a) Open the lid and remove the detector.
b) Verify the firmware version. For more information, refer to the
appropriate detector operator manual.