GasAlertMicro 5/PID/IR
User Manual
34
Initialize
A corresponding list of toxic sensors
displays. A checkbox displays beside
the current toxic sensor.
Note
Toxic 1:
List includes the PID and CO
2
sensors.
Toxic 2:
List includes the H
2
S/CO COSH
sensor.
3.
Press
G
or
H
to scroll to the new
sensor and press
C
to confirm. A
checkbox displays beside the new
sensor. To reconfigure, exit the user
options menu.
The following screen displays. The
detector deactivates and immediately
reactivates. It performs the
reconfiguration during the startup.
The new sensor must be calibrated.
Until the new sensor is calibrated, the
due date displays as
OL
(over limit)
when in normal operation.
4.
Calibrate the new sensor
immediately.
The
Initialize
option restores the origi-
nal factory default settings of the detec-
tor.
1.
Press
H
to scroll to
Initialize
and
press
C
within 20 seconds to
confirm.
2.
From the
Initialize?
screen
perform one of the following options
within 5 seconds.
•
Press
A
No
to exit, or
•
Press
C
Yes
to initialize.
If
No
is selected, the following screen
displays and the detector exits the initial-
ize option.
Summary of Contents for GasAlertMicro 5IR
Page 1: ......
Page 8: ...GasAlertMicro 5 PID IR User Manual Title Page ...
Page 10: ...GasAlertMicro 5 PID IR User Manual ...
Page 12: ...GasAlertMicro 5 PID IR User Manual ...
Page 104: ...GasAlertMicro 5 PID IR User Manual ...
Page 105: ...iERP 128900 D5615 5 English BW Technologies 2009 All rights reserved ...