33
GasAlertMicro 5/PID/IR
User Options Menu
In the following order, press and continue to hold each button until
Tech mode
displays below the
Language
option.
1. Press and hold
H
with the right index finger.
2. Press and hold
G
with the right middle finger.
3. Press and hold
C
with the left thumb.
4. Press
C
to enter
Tech mode
. The options are as
follows:
•
Sensors
•
Initialize
•
Force Calibration
(Force calibration)
•
Bump Daily
(Bump test daily)
•
Stealth Mode
•
Sleep Mode
Sensors
a
Caution
To reconfigure the sensor type, physically change the
sensor prior to entering Tech mode.
When a toxic sensor is physically removed and replaced by another toxic
sensor, the detector must be reconfigured to recognize the change.
Note
If a sensor is replaced, the detector will classify the sensor as
overdue for calibration. Calibrate the new sensor immediately.
1.
Press
H
to scroll to
Sensors
. Press
C
within 20 seconds to confirm and
access the toxic sensor menu.
2.
Press
G
or
H
to scroll to
Toxic 1
or
Toxic
2
and press
C
within
20 seconds to confirm.