Honeywell
CD3000M-1PH from 125A to 700A
User’s Manual
www.honeywell.com/imc
52
10. Maintenance
10.1
Trouble Shooting
Small problems sometimes can be solved locally with the help of the below tab of trouble shooting.
If you don’t succeed, contact us or your nearest distributor.
Symptom
Indication on
front unit
Possible reasons of the symptom
Actions
Green LED is
always light
off.
-
No voltage auxiliary power
supply to terminals 1-3 (see
wiring diagram).
-
Give auxiliary voltage supply
to terminals 1-3.
Green LED
(PW) light on
and green LED
(ON) in light off
condition.
-
No input signal.
-
Reversed polarities of input
signal.
-
Reset contact in open condition
(see wiring diagram).
-
Provide to give input signal.
-
Reverse the input signal
polarity.
-
Make link on reset terminals.
Thyristor unit
doesn’t go in
conduction with
input signal.
Green LED
(PW) in light
on condition
and green LED
(ON) in light on
condition.
-
Fuse failure.
-
Load failure.
-
Load connection interruption.
-
Thyristor faulty and always in
open circuit.
-
With HB option the yellow led
(HB) in light on condition.
-
Substitute the fuse.
-
Repair the load.
-
Provide to repair the wiring.
-
Substitute the faulty
thyristor.
-
Check the load.
Load current
flows also with
no input signal.
Green LED
(ON) always in
light off
condition.
-
Short circuit on thyristor. If
there is HB circuit the red LED
(SC) is light on.
-
Substitute the thyristor.
Check that load is not in
short circuit.
Current flows at
nominal value
but yellow LED
(HB) in light on
condition.
Yellow HB in
light on
condition.
-
HB circuit not tuned.
-
Current transformers not
properly wired (if are external
to CD3000).
-
Push CAL button in front unit
to start HB calibration
procedure.
-
Control current transformers
wiring and Push CAL button
in front unit to start HB
calibration procedure.
Red LED (SC) is
lighted also if
current is at
nominal value.
Red LED in
light on
condition.
-
HB circuit not properly tuned.
-
Push CAL button in front unit
to start HB calibration
procedure.
Thyristor unit
doesn’t work
properly.
-
Wrong input signal selection.
-
Wrong input signal calibration
(out of range).
-
Auxiliary voltage supply out of
limits.
-
Control input signal setting.
-
Repeat input calibration
procedure.
-
Verify the auxiliary voltage
supply.