Turning the BW RigRat Off
At the main measurement screen, press and hold the button. A 3-second countdown to shutoff
begins. After that, another 3-second countdown to enter the menu begins. You must release the
button during the second 3-second countdown to shut off the instrument. The screen displays
“OFF” for 3 seconds, and then the instrument is off. If you release the button during the first 3-
second countdown, the instrument continues normal operation.
Note: if you continue to hold the button during the second 3-second countdown, it enters the
main menu.
Note: You cannot turn off the BW RigRat if it is in gas alarm.
Testing Alarm Indicators
Under normal-operation mode and non-alarm conditions, the backlight can be tested by
turning the instrument on. If you perform a bump test, the buzzers and LEDs are tested.
IMPORTANT!
If any alarm does not respond, check alarm settings to make sure all alarms are enabled. If any
alarms are enabled but not functional, the instrument should not be used.
BW RigRat
66
User Manual