Supported Sensors
Gas Type
Range
Resolution
Temperature Range
COMBUSTIBLE GASES (IR-LP)
0-100%LEL
1%LEL
-40 to +140° F /
-40 to +60° C
COMBUSTIBLE GASES (IR-LP)
0-100%VOL
0.1%VOL
-40 to +140° F /
-40 to +60° C
COMBUSTIBLE GASES (IR)
0-100%LEL
1%LEL
-4 to +131° F /
-20 to +50° C
COMBUSTIBLE GASES (IR)
0-100%VOL
0.1%VOL
-4 to +131° F /
-20 to +50° C
COMBUSTIBLE GASES (LEL)
0-100%LEL
1%LEL
-40 to +140° F /
-40 to +60° C
CARBON DIOXIDE (CO
2
)
0-5%VOL/0-
50000ppm
0.01%VOL/100ppm
-4 to +131° F /
-20 to +50° C
CARBON DIOXIDE (CO
2
) (EXT.
RANGE)
0-100%VOL
0.1%VOL
-4 to +131° F /
-20 to +50° C
VOLATILE ORGANIC
COMPOUNDS (VOC)
0-2000ppm
0.1ppm
-4 to +131° F /
-20 to +50° C
CARBON MONOXIDE (CO)
0-500ppm
1ppm
-40 to +140° F /
-40 to +60° C
CARBON MONOXIDE (CO) (EXT.
RANGE)
0-2000ppm
10ppm
-4 to +131° F /
-20 to +50° C
CARBON MONOXIDE (CO) (H
2
RESISTANT)
0-2000ppm
5ppm
-4 to +131° F /
-20 to +50° C
HYDROGEN SUFLIDE (H
2
S)
0-100ppm,
0.1ppm
-40 to +140° F /
-40 to +60° C
HYDROGEN SUFLIDE (H
2
S) (EXT.
RANGE)
0-1000ppm
1ppm
-4 to +131° F /
-20 to +50° C
SULFUR DIOXIDE (SO
2
)
0-20ppm
0.1ppm
-4 to +131° F /
-20 to +50° C
OXYGEN (O
2
)
0-30%VOL
0.1%VOL
-4 to +131° F /
-20 to +50° C
NITRIC OXIDE (NO)
0-250ppm
0.5ppm
-4 to +131° F /
-20 to +50° C
NITROGEN DIOXIDE (NO
2
)
0-20ppm
0.1ppm
-4 to +131° F /
-20 to +50° C
HYDROGEN CYANIDE (HCN)
0-50ppm
0.5ppm
-4 to +131° F /
-20 to +50° C
AMMONIA (NH
3
)
0-100ppm
1ppm
-4 to +131° F /
-20 to +50° C
CHLORINE (CL
2
)
0-50ppm
0.1ppm
-4 to +131° F / -20 to
+50° C
BW RigRat
101
User Manual
Summary of Contents for BW RigRat
Page 1: ...Honeywell BW RigRat LOCAL AREA GAS MONITOR ...
Page 2: ......
Page 6: ......
Page 107: ...User Manual P N W03 4001 000 BW RigRat Revision E Honeywell September 2022 ...