510324
301C User Manual
69
4/07
Honeywell
System Programming
Tests Menu
The Tests menu provides four main options, divided between two
screens. Each of these options offers different capabilities.
The first screen presents three options:
Test sequence:
Enables you to activate each output and validate
operation of each controller keypad buttons, display
pixels and various communication protocols.
Start Sim
Sequence:
This options starts or stops the Simulation mode,
which allows a simulation of a gas concentration over
an associated scale range on all transmitters. The
simulated gas concentration values are local (on the
controller) and do not affect logging functions.
(Events will be activated for the simulation but
detection devices are not affected.)
Maximum load:
Activates all controller components
The second screen option is “Oprt Mode”, which offers three separate
operation settings: Normal, Single Tx or Debug.
Normal
Normal controller operation mode
Single Tx:
Activates the polling mode on a single transmitter.
Debug:
Activates the service mode to allow you to perform a
calibration and to test Events without triggering
actions.
8. Tests
Test sequence
Start Sim Sequence
Maximum load
8. Tests
Oprt Mode: normal
Summary of Contents for 301C
Page 1: ...Centrale Controller Manuel de l utilisateur 301C User Manual ERP 510324 4 07 ...
Page 2: ......
Page 3: ...Controller Unit 301C User Manual ERP 510324 4 07 ...
Page 4: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 14: ......
Page 108: ......
Page 110: ......
Page 114: ......
Page 115: ...Centrale 301C Manuel de l utilisateur ERP 510324 4 07 ...
Page 116: ......
Page 118: ......
Page 120: ......
Page 122: ......
Page 126: ......
Page 210: ......
Page 220: ......
Page 226: ......
Page 228: ......
Page 232: ......