510324
301C User Manual
67
4/07
Honeywell
System Programming
Network Menu
10. Press Enter to confirm or Esc to cancel. If confirmed, the controller
calibrates the sensor’s Zero. This takes only a few moments and
the display returns to the default calibration screen.
Note: Never calibrate any unit’s Zero with ambiant air. Always use
Nitrogen (N
2
) at the calibration port to calibrate the Zero.
11. To calibrate the device, scroll to Set Span** and change the span
gas calibration value using this procedure;
a.
Using the right arrow, move the cursor to xxx PPM (span
value field). Press Enter to select the field (it is editable when
flashing).
b.
Use the up or down arrows to increase or decrease the value,
press Enter to validate the new value.
c.
Move the cursor back to Set Span and press Enter to start
the calibration.
The device Span is being calibrated. The screen will display the
device’s status as “In calib...” until the calibration is complete.
**When selecting Set Span, make sure that the device has been
supplied with the appropriate calibration gas before and during the
calibration process.
Calib 001 : 301D2
Status : Normal
Set Zero
Set Span
2.60
ppm
Calib 001 : 301D2
Status : In calib...
Set Zero
Set Span
2.60
ppm
Summary of Contents for 301C
Page 1: ...Centrale Controller Manuel de l utilisateur 301C User Manual ERP 510324 4 07 ...
Page 2: ......
Page 3: ...Controller Unit 301C User Manual ERP 510324 4 07 ...
Page 4: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 14: ......
Page 108: ......
Page 110: ......
Page 114: ......
Page 115: ...Centrale 301C Manuel de l utilisateur ERP 510324 4 07 ...
Page 116: ......
Page 118: ......
Page 120: ......
Page 122: ......
Page 126: ......
Page 210: ......
Page 220: ......
Page 226: ......
Page 228: ......
Page 232: ......