7 – 6
Section 7 – Troubleshooting
ITEM
PROBLEM
SOURCE
SOLUTION
13
Invalid patient test
result.
Analyzer Internal
Controls:
Analyzer Fail/Cassette
Pass
Analyzer Internal
Controls:
Analyzer Pass/Cassette
Fail
Refer to Error/Invalid Code table.
Run TLi
IQ
QCette to verify analyzer is functioning.
Do not bump or jar analyzer during the test.
Review the Rapid fFN Cassette Kit directional insert to
ensure the correct procedure was followed.
Examine the cassette. Viscous patient samples that flow
slowly may require testing by an alternate format such as
the fFN Enzyme Immunoassay.
Examine the cassette. Cassette imperfection may cause
an invalid test result. Rerun patient sample on a fresh
cassette.
14
Computer not
present message
appears.
Data Transfer
Refer to Section 4, pp. 21, 22.