TLi
IQ
System User Manual
7 – 3
Section 7 – Troubleshooting
ITEM
PROBLEM
SOURCE
SOLUTION
5
Printer not on when
test was run.
Printer
Turn printer on. Recall the test result on the analyzer
display screen. Press PRINT/ENTER on the analyzer to
print the result.
6
Printer output in
unusual font.
Printer
Turn analyzer and printer off and back on.
7
Analyzer turned off
after calibration, or
power failure
occurred after
calibration.
Power
The calibration remains in memory. Reset calibration only
if prompted by analyzer.
8
Cassette cannot be
removed.
Analyzer
Turn analyzer off and back on to reinitialize the system.
Summary of Contents for TLi IQ
Page 1: ...User Manual System...