System User Manual
4
−
10
Section 4
−
Software Functions - Detailed Descriptions
DAILY QC
Prior to running the TLi
IQ
QCette
®
for the first time, QCette SETUP must
be performed. See page 4-29. Refer to the TLi
IQ
QCette directional insert
for more information.
Option 3 on the Main Menu screen allows the user to run the QCette.
The most recent USER ID is always displayed. Press ENTER to accept
the ID, or enter a new User ID. This field will accept 15 alpha or numeric
characters. To leave this field blank, delete the information using the
←
key.
The QCette SN must be entered to proceed to the next step. The serial
number is printed on the QCette plastic housing. The software requires
that the serial number is entered in the correct format: six numeric
characters (e.g., 001004). Enter all leading zeros.
QCette
ENTER QCette SN
>001004
ENTER-CONFIRM
QCette
USER ID
>JOHN SMITH
ENTER-CONFIRM
fFN MAIN MENU
1-TEST PATIENT
2-SET CALIBRATION
3-DAILY QC
↓