Cleaning the Exterior of the Machine

The  exterior  of  the  system  should  be  periodically  cleaned  with  a  soft
cloth, lightly moistened with water and a gentle soap. This includes the
system chassis, control panel and handles. Be sure to clean any unused
gel from the surface of the control panel after system use.

The products below have been tested and found to be compatible with
SuperSonic Imagine Aixplorer®.

Solution

Manufacturer

Cleaner/

Disinfectant

Active ingredient

Sani-Cloth Active

PDI

Cleaner

Disinfectant

Quat. Ammonia

Sani-Cloth Af3

PDI

Cleaner

Disinfectant

Quat. Ammonia

Cleaning the Monitor

A  clean,  soft  cotton  cloth  dampened  with  an  ammonia-based  window
cleaner should be used to clean the Monitor display. Do not use paper
towels,  as  these  may  scratch  the  display  monitor.  The  cleaning  agent
should be sprayed onto the cloth, not directly onto the display surface.
Take care not to get any cleaning agent into the housing of the system.
This may lead to system damage.

Cleaning the Touch Screen Display

A  clean,  soft  cotton  cloth  dampened  with  an  ammonia-based  window
cleaner  should  be  used  to  clean  the  Touch  Screen  display.  Do  not  use
paper  towels,  as  these  may  scratch  the  display  monitor.  The  cleaning
agent  should  be  sprayed  onto  the  cloth,  not  directly  onto  the  display
surface. Take care not to get any cleaning agent into the housing of the
system. This may lead to system damage.

5

Summary of Contents for SUPERSONIC IMAGINE AIXPLORER MultiWave

Page 1: ...AIXPLORER CLEANING INSTRUCTION GUIDE PM LAB 128 A 1...

Page 2: ...This document refers to the following Aixplorer User Guides and Addendums PM LAB 064 A PM LAB 026 B PM LAB 032 A 2...

Page 3: ...stem should occur at regular intervals or more frequently as needed SuperSonic Imagine provides the following recommended cleaning intervals as guidelines Part Recommended Cleaning Frequency System Di...

Page 4: ...tive Always press the On Off switch turn off the power switch at the back of the machine and disconnect the system from the wall outlet before performing maintenance or cleaning the system It is recom...

Page 5: ...tor A clean soft cotton cloth dampened with an ammonia based window cleaner should be used to clean the Monitor display Do not use paper towels as these may scratch the display monitor The cleaning ag...

Page 6: ...e bottom side of the system To remove the rear air filter open the cover on the back To remove the side air filter open the left side cover with a pen It is recommended that air filters be cleaned mon...

Page 7: ...installing WARNING Before performing maintenance or cleaning always press the On Off switch turn off the power switch at the back of the machine and then disconnect the system from the wall outlet Inc...

Page 8: ...HS 1M is IP54 rated when the cable is attached This means that liquids and dusts will not penetrate into the housing However the scanner should not be submerged in water or other liquids It is also go...

Page 9: ...Never use solvents e g acetone benzene ether or phenol based agents on the housing or window solvents may damage the finish or the window 9...

Page 10: ...ucer prior to cleaning for damage or degeneration to the lens housing cable and connector Do not use or apply liquid cleaning agents to a damaged or defective transducer Do not immerse the transducer...

Page 11: ...luids Proper cleaning and disinfection are necessary to prevent disease transmission and infection control It is the responsibility of the equipment user to verify and maintain the effectiveness of th...

Page 12: ...intended to remove all foreign matter blood tissue protein scanning gel etc from the device DISINFECTION The level of disinfection required for a device is dictated by the type of tissue it will cont...

Page 13: ...cations high level disinfection is required and only one of the high level disinfectants chosen from the list below should be used High level disinfection for semi critical is recommended carefully fo...

Page 14: ...When using an FDA or CE cleared as appropriate disinfecting agent carefully follow the instructions provided by the manufacturer of the product 3 For low level disinfection the lens and housing of th...

Page 15: ...opyl alcohol Alkylamine YES Alkazyme Alkapharm Cleaner Disinfectant Quat Ammonia YES Anioxyde 1000 Anios Disinfectant Peracetic acid YES YES Antigermix S1 Germitec High level disinfectant UV C YES Cid...

Page 16: ...x Cleaner Proteolytic enzymes YES Mikrobac Tissues Bode Disinfectant Benzyl C12 18 alkyldimethyl ammonium chlorides didecyl dimethyl ammonium chloride YES YES Milton Milton Disinfectant Sodium Hypochl...

Page 17: ...Cleaner Chlorhexidine Gluconate Isopropyl Alcohol YES Tristel Duo Tristel Solutions Limited Disinfectant Chlorine dioxide YES YES Tristel Solo Tristel Solutions Limited Disinfectant Hexamethylene bigu...

Page 18: ...to be compatible with SuperSonic Imagine Aixplorer transducers material solely Questions regarding efficacy instructions for use and proper handling should be directed to the manufacturer Please chec...

Page 19: ...XP5 1 XC6 1 SLH20 6 and SE12 3 transducers may be immersed in liquid to a level of 1m 39 37in above the cable strain relief junction See figure below According to the IEC 60529 standard all Aixplorer...

Page 20: ...20...

Page 21: ...th care facilities ANSI AAMI ST58 2005 Sterilization and Disinfection of Medical Devices General Principles Centers for Disease Control Division of Healthcare Quality Promotion http www cdc gov ncidod...

Page 22: ...22...

Reviews: