Soaking Limits
If soaking a transducer is required, the SL15-4, SL18-5, SL10-2, SC6-1,
SMC12-3 transducers may only be immersed in liquid to a level of 3cm
(1.18 in.) above the cable-strain relief junction. The SEV12-3, SLV16-5,
XP5-1, XC6-1, SLH20-6 and SE12-3 transducers may be immersed in
liquid to a level of 1m (39.37in.) above the cable-strain relief junction.
See figure below.
According to the IEC 60529 standard, all Aixplorer probes are IPX7
according to the figure below.
Never immerse the probe connector. Never immerse the cable-connector
junction. Do not leave the transducer immersed longer than required for
disinfection. See figure below.