Transducer used in semi-critical application should be cleaned and high
level disinfected after use even if a sheath was used.
Transducers used in critical applications should be sterilized and the use
of a sterile sheath is recommended. Transducer used in critical application
should be cleaned and sterilized after use even if a sheath was used.
General Cleaning and Disinfection
Procedures for the Aixplorer®
Transducers
The SuperSonic Imagine Aixplorer® transducers are intended for use
in non-critical and semi-critical imaging applications. The transducer
must be thoroughly cleaned and disinfected after each use. In general,
only low-level disinfection is required for non-critical applications of the
SuperSonic Imagine Aixplorer® transducer.
For semi-critical applications, high-level disinfection is required and only
one of the high-level disinfectants chosen from the list below should be
used.
High-level disinfection for semi-critical is recommended, carefully
following the manufacturer’s instructions with regard to application and
contact time as indicated on the product label.
PROCEDURE TO PREPARE TRANSDUCER FOR
DISINFECTION
1. After every patient exam, carefully wipe the face and housing of the
ultrasound transducer to remove any traces of ultrasound coupling gel.
2. Inspect the transducer for any signs of damage such as cracks,
splitting, delamination of the face or cable damage. If any damage is
noted, do not proceed with the cleaning or disinfecting procedures and
call your authorized SuperSonic Imagine Service representative for
assistance.
3. Wipe the transducer face, housing and cable with a water-moistened
soft cloth. The use of a mild soap, low-level cleaner or enzymatic
cleaner (from the table below) is acceptable.
13