Hoefer PS200HC User Manual Download Page 21

4   Press the SELECT button to position the indicator 

light under the Clock symbol. If desired, use the 
UP button to set the timer to a desired duration. If 
the run is continuous, set the timer value to “0”.

5  Press the START/STOP button to start the run.

Example: Set up conditions for running a semi-dry  
transfer of a standard-size PAGE gel at 0.18 A and 
30 V, with automatic crossover.

The power supply can automatically crossover between 
two different maximum settings, if desired. This will 
be illustrated below.

1   Press the CONSTANT button to position the  

indicator light under “A”.

2   Press the UP and DOWN buttons to increase the  

LED display value to 0.18.

3   Press the SELECT button to position the indicator 

light under “V”. Observe that the volt limit has 
been automatically set to the maximum value of 
200 V. Press the UP and DOWN buttons to adjust 
this value down to 30 V.

4   Press the SELECT button to position the indicator 

light under the Clock symbol. If desired, use the 
UP button to set the timer to a desired duration. If 
the run is continuous, set timer to 0.

5  Press the START/STOP button to start the run.

Both Voltage and Current are monitored during 
the run. The power supply initially outputs 0.18 A 
constant current. The voltage output will be 
<30 V, but may rise during the transfer run, due 
to gradual depletion of the buffer during the run. 
If the voltage never exceeds 30 V during the run, 
then the gels will run at 0.18 A for the entire 
run. If the voltage increases to 30 V, then the 
power supply will automatically “crossover” to 
constant voltage mode, in order to keep voltage 
from exceeding 30 V. Now, output current will 
slowly decrease below the original 0.18 A setting 
because 30 V is now the limiting value in the 
curcuit.

 p11

Note: 

If this crossover condi-

tion occurs and is not desired, 
change the voltage or current 
setting to the power supply’s 
operational maximum (2.0 A or 
200 V).

 

Summary of Contents for PS200HC

Page 1: ...user manual um PS200HC IM Rev C0 07 12 PS200HC Power Supply...

Page 2: ...formation viii Power supply function and description 1 Unpacking 3 Specifications 4 Parameter overview 5 Power supply operation 7 Connections 8 Operation 9 Power Failure Detection PFD 12 Troubleshooti...

Page 3: ...it n pou it v laborato i pouze Pouze p slu enstv a sti schv len nebo poskyt nut ch Hoefer Inc mohou b t pou ity pro provoz dr bu a dr b tohoto v robku Pozor Proto e tento n stroj m e vyvinout dostate...

Page 4: ...spanning zou in overeenstemming met EN61010 1 2001 moeten zijn Houd het instrument zo droge en schone zoals mogelijk Bij Wis regelmatig met een zacht temper doek Verhuur het instrument droogt volledig...

Page 5: ...sr stung zur Verf gung gestellte Schutz verschlechtert werden Dieses Instrument wird f r den Innenlaborgebrauch nur daf r entworfen Nur Zus tze und Teile genehmigten oder lieferten durch Hoefer Inc ka...

Page 6: ...ig sjokk m bli vd bekymring i dets drift Dette instrumentet er utformet i samsvar med EN61010 1 2001 elektrisk sikkerhetsstandard Likevel burde bli brukt det bare av skikkelig utdannede oper at rer Le...

Page 7: ...o h mido macio Deixe o instrumento secar completamente antes de uso N o opera o instrumento em humidade extrema acima de 80 Evite condensa o deixando o equili brate de unidade a temperatura ambiental...

Page 8: ...as bara av riktigt utbildade operat rer L s denna hela handbok f re anv nda instrumentet och anv nd bara enligt under visningarna Instrumentet m ste alltid anv ndas med jorden blyet av kraften repet r...

Page 9: ...e Kontakt mit einem autorisierten Beauftragten des Herstellers auf um Informationen hinsichtlich der Entsorgung Ihres Ger tes zu erhalten Questo simbolo indica che i rifiuti derivanti da apparecchiatu...

Page 10: ...CE certification This product meets the requirements of applicable CE directives A copy of the corresponding Declaration of Conformity is available on request The CE symbol and corresponding declarati...

Page 11: ...current status Param eters are set by pressing UP DOWN arrow buttons in combination with a SELECT button Four pairs of output jacks accept 4 mm plugs which operate in parallel The output jacks are col...

Page 12: ...p2 Fig 1 Front view of PS200HC Power Supply 3 digit display UP DOWN arrows SELECT display highlighted parameter CONSTANT V or A toggle START STOP polarity Output jacks polarity...

Page 13: ...er Inc sales office Inspect all components for damage that may have occurred while the unit was in transit If any part appears damaged contact the carrier immediately Be sure to keep all packing mater...

Page 14: ...lel Output connector 4 mm plugs Timer 1 999 min 1 min increment Environmental operating Indoor use 4 40 C conditions Relative humidity 0 80 noncondensing Altitude 2000 m Pollution degree 2 Installatio...

Page 15: ...d or power requirement does not exceed the power supply rating Crossover occurs when the second parameter limit is reached as system resistance changes Fig 3 At this point the second parameter becomes...

Page 16: ...ower When this occurs both the constant V and the constant A lights will be lit at the same time p6 current Voltage limit Crossover from constant A to constant V Current limit resistance increasing ov...

Page 17: ...ng or discon necting the voltage leads 2 Keep fingers away from metal components when connecting or disconnecting the voltage leads p7 WARNING Use only one hand when making or breaking a connection to...

Page 18: ...e power supply so that the fan on the back panel and the vents on the bottom are not obstructed 2 Connect the HV power leads Connect the apparatus to the power supply by plugging the lead connectors i...

Page 19: ...s to set it to the desired value The power supply can then operate in Crossover mode as explained on page 5 4 Set the timer optional Press the SELECT button to move the green indicator light to under...

Page 20: ...ng the CONSTANT button will reset the power supply Example Set up conditions for running a standard transfer tank e g TE62 at a constant current of 1 0 A 1 Press the CONSTANT button to position the in...

Page 21: ...his value down to 30 V 4 Press the SELECT button to position the indicator light under the Clock symbol If desired use the UP button to set the timer to a desired duration If the run is continuous set...

Page 22: ...The time shown at the end of the run is the time the gel ran since power was restored not the total time of the run The default settings for Power Failure Detection are Run Time Setting PFD Code Time...

Page 23: ...Check buffer volumes Check all voltage leads Check the fit of power supply adapters if used The output current may have fallen below the 01 A limit of the power supply Press any button to clear the er...

Page 24: ...el Use a screwdriver to pry out the fuse compartment There are two fuses in the compartment Check and replace any blown fuses with a fuse of the same size and ratings FUSES 2 250 V 5 0 A 5 20 mm fast...

Page 25: ...l Hoefer Inc for advice before using Technical service and repair Hoefer Inc offers complete technical support for all of our products If you have any ques tions about how to use this product or would...

Page 26: ...lliston MA 01746 Toll Free 1 800 227 4750 Phone 1 508 893 8999 Fax 1 508 893 0176 E mail support hoeferinc com Web www hoeferinc com Hoefer is a registered trademark of Hoefer Inc 2012 Hoefer Inc All...

Reviews: