Parameter overview

Power parameters

Because the power output limit is 200 W, and  
W = V × A, the maximum current of 2.0 A 
can be delivered only if voltage is 

100 V. 

Conversely, the maximum voltage of 200 V can 
be delivered only if the current is 

1.0 A.

Programmed limits

Variable 

Range 

Increment

Voltage 

5 – 200 V 

1 volt

Current 

0.01– 2.00 A 

0.01 amp

Time 

0 – 999 minutes 

1 minute

Timer = 0 = continuous run
Timer = 1   –   999 = timed run in minutes

 
Power supply output limits

The PS200HC can be programmed to limit volt-
age (V) or current (A). The power supply will 
operate at the set limit provided the “load,” or 
power requirement, does not exceed the power 
supply rating.

“Crossover” occurs when the second parameter 
limit is reached as system resistance changes 
(Fig 3). At this point, the second parameter 
becomes constant, and the previously unchang-
ing param eter adjusts downward as system 
resistance changes. (System resistance is affected 
by discontinuous buffers, changes in tempera-
ture, and buffer depletion.) When the power 
supply is delivering power, the light above the 
CONSTANT button indicates which parameter 
is being held constant.

 p5

Please note the following power 
output limitations:

•  The maximum current of 

2.0 A is available at 100 V 
or less. 

•  The maximum voltage setting 

of 200 V is available at 1.0 A 
and less.

Note:  In most cases, only one 
parameter can be held constant 
at a time.

Summary of Contents for PS200HC

Page 1: ...user manual um PS200HC IM Rev C0 07 12 PS200HC Power Supply...

Page 2: ...formation viii Power supply function and description 1 Unpacking 3 Specifications 4 Parameter overview 5 Power supply operation 7 Connections 8 Operation 9 Power Failure Detection PFD 12 Troubleshooti...

Page 3: ...it n pou it v laborato i pouze Pouze p slu enstv a sti schv len nebo poskyt nut ch Hoefer Inc mohou b t pou ity pro provoz dr bu a dr b tohoto v robku Pozor Proto e tento n stroj m e vyvinout dostate...

Page 4: ...spanning zou in overeenstemming met EN61010 1 2001 moeten zijn Houd het instrument zo droge en schone zoals mogelijk Bij Wis regelmatig met een zacht temper doek Verhuur het instrument droogt volledig...

Page 5: ...sr stung zur Verf gung gestellte Schutz verschlechtert werden Dieses Instrument wird f r den Innenlaborgebrauch nur daf r entworfen Nur Zus tze und Teile genehmigten oder lieferten durch Hoefer Inc ka...

Page 6: ...ig sjokk m bli vd bekymring i dets drift Dette instrumentet er utformet i samsvar med EN61010 1 2001 elektrisk sikkerhetsstandard Likevel burde bli brukt det bare av skikkelig utdannede oper at rer Le...

Page 7: ...o h mido macio Deixe o instrumento secar completamente antes de uso N o opera o instrumento em humidade extrema acima de 80 Evite condensa o deixando o equili brate de unidade a temperatura ambiental...

Page 8: ...as bara av riktigt utbildade operat rer L s denna hela handbok f re anv nda instrumentet och anv nd bara enligt under visningarna Instrumentet m ste alltid anv ndas med jorden blyet av kraften repet r...

Page 9: ...e Kontakt mit einem autorisierten Beauftragten des Herstellers auf um Informationen hinsichtlich der Entsorgung Ihres Ger tes zu erhalten Questo simbolo indica che i rifiuti derivanti da apparecchiatu...

Page 10: ...CE certification This product meets the requirements of applicable CE directives A copy of the corresponding Declaration of Conformity is available on request The CE symbol and corresponding declarati...

Page 11: ...current status Param eters are set by pressing UP DOWN arrow buttons in combination with a SELECT button Four pairs of output jacks accept 4 mm plugs which operate in parallel The output jacks are col...

Page 12: ...p2 Fig 1 Front view of PS200HC Power Supply 3 digit display UP DOWN arrows SELECT display highlighted parameter CONSTANT V or A toggle START STOP polarity Output jacks polarity...

Page 13: ...er Inc sales office Inspect all components for damage that may have occurred while the unit was in transit If any part appears damaged contact the carrier immediately Be sure to keep all packing mater...

Page 14: ...lel Output connector 4 mm plugs Timer 1 999 min 1 min increment Environmental operating Indoor use 4 40 C conditions Relative humidity 0 80 noncondensing Altitude 2000 m Pollution degree 2 Installatio...

Page 15: ...d or power requirement does not exceed the power supply rating Crossover occurs when the second parameter limit is reached as system resistance changes Fig 3 At this point the second parameter becomes...

Page 16: ...ower When this occurs both the constant V and the constant A lights will be lit at the same time p6 current Voltage limit Crossover from constant A to constant V Current limit resistance increasing ov...

Page 17: ...ng or discon necting the voltage leads 2 Keep fingers away from metal components when connecting or disconnecting the voltage leads p7 WARNING Use only one hand when making or breaking a connection to...

Page 18: ...e power supply so that the fan on the back panel and the vents on the bottom are not obstructed 2 Connect the HV power leads Connect the apparatus to the power supply by plugging the lead connectors i...

Page 19: ...s to set it to the desired value The power supply can then operate in Crossover mode as explained on page 5 4 Set the timer optional Press the SELECT button to move the green indicator light to under...

Page 20: ...ng the CONSTANT button will reset the power supply Example Set up conditions for running a standard transfer tank e g TE62 at a constant current of 1 0 A 1 Press the CONSTANT button to position the in...

Page 21: ...his value down to 30 V 4 Press the SELECT button to position the indicator light under the Clock symbol If desired use the UP button to set the timer to a desired duration If the run is continuous set...

Page 22: ...The time shown at the end of the run is the time the gel ran since power was restored not the total time of the run The default settings for Power Failure Detection are Run Time Setting PFD Code Time...

Page 23: ...Check buffer volumes Check all voltage leads Check the fit of power supply adapters if used The output current may have fallen below the 01 A limit of the power supply Press any button to clear the er...

Page 24: ...el Use a screwdriver to pry out the fuse compartment There are two fuses in the compartment Check and replace any blown fuses with a fuse of the same size and ratings FUSES 2 250 V 5 0 A 5 20 mm fast...

Page 25: ...l Hoefer Inc for advice before using Technical service and repair Hoefer Inc offers complete technical support for all of our products If you have any ques tions about how to use this product or would...

Page 26: ...lliston MA 01746 Toll Free 1 800 227 4750 Phone 1 508 893 8999 Fax 1 508 893 0176 E mail support hoeferinc com Web www hoeferinc com Hoefer is a registered trademark of Hoefer Inc 2012 Hoefer Inc All...

Reviews: