-15-
Q1E-EP0428
5.1.
Point of use (Pre-cleaning)
Pre-cleaning should be done immediately after each use. The procedure
is as follows:
A)
EUP-C514 probe
1)
Remove the protective cover.
2)
Clean the probe of all patient
’
s blood or fluid with running tap
water until the surface of the probe looks visually clean.
3)
Wipe the whole surface of the probe with gauze pad and remove
superficial visible impurities.
B)
Magnetic sensor Attachment
1)
Remove the Attachment and the magnetic sensor from the probe.
2)
Immerse the Attachment in sufficient amount of high quality tap water.
Scrub it using soft cloth to remove all visible soil and from its
surface.
5.2.
Containment and transportation
Putting the contaminated equipment into exclusive shock and damage proof
container for transportation is recommended. It is recommended that
instruments are reprocessed as soon as possible and not later than 4
hours after usage.
5.3.
Manual Cleaning and disinfection
a)
Prepare following items before manual
cleaning and disinfection:
b)
Detergent: Cidezyme
®
(Johnson & Johnson,
#2258) or another cleaning agent with
approved material compatibility for this
medical device.
c)
Disinfectant: Cidex
®
OPA (Johnson &
Johnson, # 20391) or another disinfectant
with approved material compatibility for
this medical device.
d)
Two tanks, one for cleaning and one for disinfection - optional:
1 additional tank for rinsing with deionized/tap water (sufficient
size for immersion of the submergible part of the probe at full
length)
e)
Soft, fluff free cloth or single use towel
f)
Personal protective equipment (gloves, water repellent protective
skirt, face protection mask or protective glasses, see also
instructions of the manufacturer for the detergent and the
disinfectant)
Point of use
(Pre-cleaning)
Containment and
transportation
Manual Cleaning
Rinsing after manual
cleaning
Manual Disinfection
Rinsing after manual
disinfection
Manual cleaning and
disinfection