4.
Option of Hitachi ultrasound diagnostic sensor
In case of using RVS (Real-time Virtual Sonography), confirm that type
of the Magnetic sensor. There are two types of magnetic sensors for
the EUP-C514, the EZU-RV2S and the EZU-RV3S.
The Magnetic sensor (EZU-RV2S) and the Magnetic sensor (EZU-RV3S) are
shown in Fig. 3 and Fig.8.
The use of the EUP-C514 with either of the magnetic sensors enables
the user to perform RVS (Real-time Virtual Sonography).
4.1.
Magnetic Sensor (EZU-RV2S)
The Magnetic sensor (EZU-RV2S) as shown in Fig.3 is a magnetic sensor
for the EUP-C514.
4.1.1.
How to attach the Magnetic Sensor
The Procedure of attaching the Magnetic sensor is as follows.
1)
Confirm that the probe, the Magnetic sensor attachment and the Spacer
for EZU-RV2S are disinfected or sterilized.
2)
Connect the probe, operate the ultrasound diagnostic scanner, and
adjust the image according to the instructions given in the operation
manual for the ultrasound diagnostic scanner.
3)
To use Real-time Virtual Sonography (RVS), attach the Magnetic sensor
as shown below.
Fig. 3 Magnetic sensor (EZU-RV2S)
Magnetic sensor
(black)
Projected part