38
Query Automatic detection Line Frequency
Syntax
Query
:SYSTem:LFRequency:AUTO?
Response
<ERROR/50/60>
Description
ERROR
…
Detection error
50
…
50Hz
60
…
60Hz
Example
:SYSTem:LFRequency:AUTO?
60
Detected 60Hz
(31) Reset
Initialize Instrument (System Reset)
Syntax
command
:RESet
<SYSTem/NORMal>
Description
SYSTem … Initializes all data incluse panel data settings
NORMal … Initializes all data except panel data settings
Example
:RESet SYSTem
Execute system reset.
(32)
EXT I/O
Set and Query Trigger edge
Syntax
command
:IO:EDGE
<ON/OFF>
Query
:IO:EDGE?
Response
<ON/OFF>
Description
Set trigger edge
ON … PNP
:
fall
、
NPN
:
rise
OFF … PNP
:
rise
、
NPN
:
fall
Example
:IO:EDGE
ON
Set trigger edge to ON edge
:IO:EDGE?
ON
Trigger edge has been ON edge
Set and Query TRIG Signal's Filter Function
Syntax
command
:IO:FILTer:STATe
<ON/OFF>
Query
:IO:FILTer:STATe?
Response
<ON/OFF>
Description
Set to use the trigger filter.
ON … Use trigger filter.
OFF … Not use trigger filter.
Example
:IO:FILTer:STATe
ON
Set trigger filter to ON
:IO:FILTer:STATe?
ON
Trigger filter has been set to ON.