29
Query contact check cable length
auto
Syntax
Query
:CONTactcheck:CABLe:AUTO?
Response
<0
~
1(NR1)>
Description
0 … cable length auto OFF
1 … cable length auto ON
Example
:CONTactcheck:CABLe:AUTO?
1
Contact check calbe length auto has been ON
Set and Query contact check delay
Syntax
command
:CONTactcheck:DELay
<delay time(s)(NR2)>
Query
:CONTactcheck:DELay?
Response
< delay time(s)(NR2)>
Description
<delay time(s)(NR2)> = 0.0
~
9.999
Example
:CONTactcheck:DELay 1.23
Set delay time to 1.23 seconds
:CONTactcheck:DELay?
1.230
Delay time has been 1.23 seconds
Query Jig capacity open correction
Syntax
Query
:OPEN?
Response
<0
~
1(NR1)>
Description
Execute jig capacity open, return the judgnent
0 … Fail (NG)
1 …Success (OK)
Example
:OPEN?
1
Success jig capacity open
Query Jig capacity open value
Syntax
Query
:OPEN:VALue?
Response
<open value(F)(NR3)>
Description
Return jig open value
Unexecuted … 99.999E-99
Over … 99.999E-12
Example
:OPEN:VALue?
1.412E-12
Jig open value has been 1.412pF
Summary of Contents for Super megohm SM7110