23
Set and Query
average hold times
Syntax
command
:AVERage:COUNt
<2
~
255(NR1)>
Query
:AVERage:COUNt?
Response
<2
~
255(NR1)>
Example
:AVERage:COUNt 5
Set average hold time to 5
:AVERage:COUNt?
5
Average hold time has been 5
(13)
Sequence Program function
Set and Query Sequence Program
Syntax
command
:SEQuence:STATe
<ON/OFF>
Query
:SEQuence:STATe?
Response
<ON/OFF>
Example
:SEQuence:STATe
ON
Set sequence program to ON
:SEQuence:STATe?
ON
Sequence program has been ON
Set and Query Sequence program number
Syntax
command
:SEQuence:NUMBer
<0
~
9(NR1)>
Query
:SEQuence:NUMBer?
Response
<0
~
9(NR1)>
Example
:SEQuence:NUMBer 2
Set Sequence program number to 2
:SEQuence:NUMBer?
2
Sequence program number has been 2
Set and Query Sequence time
Syntax
command
:SEQuence:TIME
<sequence number(NR1)>,
<discharge time(s)(NR2)>,
<charge time(s) (NR2)>,
<measurement time(s) (NR2)>,
< discharge time (s) (NR2)>
Query
:SEQuence:TIME?
< sequence number (NR1)>
Response
< sequence number (NR1)>,< discharge time (s) (NR2)>,
< charge time (s) (NR2)>,< measurement time (s) (NR2)>,
< discharge time (s) (NR2)>
Description
< sequence number (NR1)> = 0
~
9
< discharge time (s)(NR2)> = 0.000
~
999.999
< charge time (s) (NR2)> = 0.001
~
999.999
< measurement time (s) (NR2)> = 0.00
~
999.999
< discharge time (s) (NR2)> = 0.000
~
999.999
Example
:SEQuence:TIME 1,0.1,0.5,3.0,4.5
Set sequence 1
discharge time = 0.1 seconds
charge time = 0.5 seconds
measurement time = 3.0 seconds
discharge time = 4.5 seconds
:SEQuence:TIME? 1
1,0.100,0.500,3.000,4.500
sequence 1 has been upper times