27
6.5 Comparison and Judgment Function
This function is provided to sound a built-in buzzer, and turn on the
COMP OUT G/+ te rminals when the measured insulation resistance is
lower than the preset judgment value and make the COMP OUT G/+
terminals are of open collector as shown as an equivalent circuit in Fig.
6.5.3. Use this circuit within the conditions shown below:
Voltage: 50 V or less
Current: 50 mA or less
Fig. 6 .5 .1 shows a circuit d iag ram for connection with the COMP OUT
terminals.
Fig. 6.5.2 shows the external input/output terminal block, including the
COMP OUT G/+ terminals.
Fig . 6 .5 .1
Fig. 6.5.2 Fig. 6.5.3
G
+
COMPOUT