5.4 Check the Measuring Rods
1) Confirm that V.OUT indicator turns off and that the Rx
/+
measuring
terminals don’t carry a me asuring voltage across them.
2) Plug the accessory measuring rods to the Rx
+
/
me asuring terminals
until they are stopped. (Red measuring rod to the Rx
me asuring
terminal / Black measuring rod to the Rx
+
me asuring terminal)
3) Set the RANGE selector switch to the
×
1 position – the minimum Multiplier.
4) Set the VOLTAGE SELECTOR switches to the lowest voltage available.
5)
Connect the measuring rods
each other. Do not place the measuring rods
4)
anywhere .
6) Push MEASURE / DISCHARGE Switch.
7)
Confirm that the analog indicating needle is shown leftward and that the
6 )
indicated value is shown blinking.
8)
Push MEASURE / DISCHARGE Switch. Confirm that
the Rx
/+
measuring
7)
terminals don’t carry a measuring voltage across them.
9)
Separate the measuring rods
.