![Hioki SM-8200 Series Instruction Manual Download Page 28](http://html.mh-extra.com/html/hioki/sm-8200-series/sm-8200-series_instruction-manual_2137370028.webp)
16
⑯
External Input/Output Terminal Block: This terminal block carries
signals measured result judgment outputs, optional DC output, and
remote signal input.
For details, see the following sections;
6.5 Comparison and Judgment Function
8. MEMORY CONTROLLED MEASUREMENT
9.2 DC Signal Output
⑰
HV-EN (High Voltage Enable) Interlocking Connector: This connector is
provided for connection of an external interlocking switch on the
measuring jig designed to block the application of a high voltage
across the Rx
+/
measuring terminals
④
for safety of the operator
when the switch is in the off position.
If such a switch is not provided, keep this connector plugged with
the accessory shorting plug, instead.
⑱
RS-232C Interface Connector: This connector is provided for RS-232C