Replacing Fuses
6.4
Replacing Fuses
If a fuse is blown, replace it with a new one as follows.
WARNING
Replace the fuse only with one of the speci
fi
ed type,
characteristics, rated current, and rated voltage.
Do not use fuses other than those speci
fi
ed (especially,
do not use a fuse with higher-rated current) or do not
short circuit and use the fuse holder. Doing so may
damage the instrument and result in personal injury.
Speci
fi
ed fuses
Rating
Speci
fi
cations
For
μ
A/mA terminal
(DT4253)
250 mA/
1000 V
Manufacturer: HOLLYLAND
Breaking characteristic: Fast-blow type
Breaking capacity:
50 kA AC/ 30 kA DC
Size:
10.3 mm × 38 mm
For A terminal
(DT4252)
11 A/
1000 V
The fuses can be purchased via authorized Hioki distributor or
reseller.
When removing the fuse, do not apply excessive force on the fuse
holder. If the fuse holder is deformed, the connection becomes
poor and the instrument cannot measure the current.