![Hioki DT4251 Instruction Manual Download Page 53](http://html1.mh-extra.com/html/hioki/dt4251/dt4251_instruction-manual_2137396053.webp)
47
Measuring Resistance
3.6
Measuring Resistance
Resistance is measured.
To measure the low resistance accurately, it is necessary to cancel
the resistance of the test leads. Perform zero adjustment for the
displayed value using the relative value display (relative function) in
advance.
WARNING
Before measuring, be sure to turn off the power to the
measurement circuit. Otherwise, electric shock may
occur or the instrument may be damaged.
2
2
33
Red
Black
1
1
The open terminal voltage is approx. 1.8 V or less. The
measurement current (DC) varies depending on the range.
To avoid damage to the measurement object, check the
speci
fi
cations before use.
Summary of Contents for DT4251
Page 2: ......
Page 6: ...iv ...
Page 20: ...14 Usage Notes ...
Page 42: ...36 Using the Carrying Case ...
Page 100: ...94 Replacing Fuses ...
Page 102: ...Appx 2 RMS and Average ...
Page 103: ...13 09 ...
Page 104: ......
Page 105: ......
Page 106: ......