45
Checking Continuity
3.4
Checking Continuity
The input short circuit is detected and informed via a buzzer and red
LED.
WARNING
Before measuring, be sure to turn off the power to the
measurement circuit. Otherwise, electric shock may
occur or the instrument may be damaged.
2
2
33
Red
Black
1
1
Detection
Threshold
Buzzer
Red LED
Short circuit detection
25
Ω
± 10
Ω
Sounds
(continuous
buzzer sound)
Turns on
Open detection
245
Ω
± 10
Ω
Does not sound
Turns off
A buzzer sounds before the red LED lights up.