4.2 Toggling the Percentage Display (4 - 20mA/0 - 20mA)
When measuring with the DCmA function, press the
DUAL
key three times to display the percentage display in
the sub display.
This procedure sets the percentage display method.
4.2 Toggling the Percentage Display
(4 - 20mA/0 - 20mA)
4-20 mA
Converts the current over a range of 4 mA to 20
mA into a percentage from 0 to 100%.
0-20 mA
Converts the current over a range of 0 mA to 20
mA into a percentage from 0 to 100%.
3
4
1. In order to display the Power On
Option setting screen, hold down
the
SHIFT
key while turning the
function switch.
2. Press the
HOLD
key 5 times to dis-
play the Percentage Display set-
ting screen.
3. Press the
RANGE
key or the
DUAL
key to select the percentage dis-
play setting.
4. Confirm the selection by pressing
the
Hz
key.
5. Hold down the
SHIFT
key to enter
the measurement screen.
Otherwise, turn the function switch
to OFF to finish setup.
4-20 mA
(default)
/ 20 mA
1
SHIFT +
Power On
2
(5 times)
5
Select through
power on option
Percentage
conversion value =
Measured value[mA] - 4[mA]
16[mA]
X 100[%]
Measured value[mA]
20[mA]
X 100[%]
Percentage
conversion value =