Transducer Replacement
Turn off the device, remove the Nebulization chamber (9) and take off the lid of
nebulization chamber. Undo the chromium-plated nut underneath the nebulization
chamber and remove the complete transducer assembly (41). Replace the new
transducer so that the shiny side is facing towards the nebulization chamber. Reapply
and tighten the chromium nut.
Note
The device will not operate if the transducer is not installed correctly. Also, whenever the
transducer has been replaced and before the chamber is connected in place, fill the
chamber with water and check for possible leaks.
Chromium-plated
Nut, transducer
installed
Transducer
Chromium-plated
nut
10. Cleaning, Disinfection, Sterilization
Cleaning
The device and 5 feet mobile stand can be cleaned with a mild detergent and damp
cloth. For special cleaning/disinfecting please follow the manufacturers cleaning
instructions detailed in this user manual.
NO parts should be exposed to excessive pressure during the cleaning process.
Pay specific attention to the transducer surface ensuring it is clean. Clean lime scale
residue by wiping, not scrubbing. Abrasive scrubbing can damage the transducer.