User Manual
Test
Level
Ambient
TS EN 61000-4-8
Main frequency magnetic
area immunity
30 mA
Its power frequency magnetic area
should be in hospital or mains quality.
Our devices have no magnetic break
effect on power network.
TS EN 61000-4-11
Voltage Dips, Short
Interruptions and Voltage
Changes Immunity
%0- 0,5 time
0°-45°-90°-135°-180°-
225°270°-315°
%0 – 1 time
%70- 25/30 times
0° single phase
Power feeding should be on mains or
hospital installation quality, in building
power input, Leakage Current Protective
Relay should be plugged if there is not.
If a user wants to use the device in
electric cuts, it is required to use
Uninterruptable Power Supply.
TS EN 55011
Terminals disruptive
Voltage Emission Break
Class B
RF transmission is at low level, it does
not jam in other devices.
TS EN 61000-3-2
Harmonics
TS EN 61000-3-3
Voltage Fluctuations and
Flicking
Class A
It is suitable to be operated at Low
Voltages in mains and hospital
installation type, it can directly be
connected to a socket. If Instantaneous
Voltage changes occur often, Voltage
Regulator is recommended.
Distances recommended for preventing RF frequencies
Placing the devices within the following distances from fixed and portable RF devices
will assist in preventing interferences.
Nominal power of
the transmitter
150KHz-80MHz ISM bands
and 80MHz-800MHz
u=0.5
√
G
800MHz-2.5GHz
u=2.3
√
G
0.01 W
0.12 m
0.24 m
0.1 W
0.37 m
0.74 m
1 W
1.17 m
2.34 m
10 W
3.69 m
7.38 m
100 W
11.67 m
23.34 m