Electric shock! Before replacing the fuses, disconnect the device from the mains.
The main fuse (2A) is located inside the IEC socket on the back of the device. Remove
the power cord from the device. To remove fuse holder, use a small flat screwdriver
pulling the holder forward. Replace fuse with the same fuse rating and replace fuse
holder by pushing back into position.
8a. Replacement of the main fuse.
Caution
9. Maintenance, Repair and Disposal
Looking after your device and keeping it in good condition will ensure the device lasts
the expected service life. To keep the device in good working order, we advise to clean
the device after every use, and would suggest functional testing every six months as
per the instructions detailed in this user manual. Should the device ever become faulty,
technical support maybe required.
Bacteria filters can be disposed of with normal domestic waste. Nebulization chamber,
mobile trolley wheels, aerosol tubing which include PC, ABS and PVC can also be
considered as normal waste. The outer casing of the device itself is constructed from
aluminium and would need to be recycled or disposed of under WEEE regulations.
Devices returned to the manufacturer for disposal, will be processed under WEEE
Regulations.
Bacteria Filter Replacement
Using the approved bacteria filter (19) is mandatory to prevent contamination during
aerosol therapy. The filter should be replaced after 48 hours of continuous use, or
1 week for non-continuous use. Low aerosol output is also a good indication of a
contaminated/occluded filter.