Removing grease deposits
NOTICE
Possible damage to the calibration sphere when cleaning
The use of solvents or cleaners is
not
recommended, as they may damage the calibration sphere.
Light grease deposits can be removed with extreme care using a clean, soft, lint-free micro-fibre cloth and
isopropyl alcohol.
However, any such cleaning is performed entirely at own risk.
1. Moisten a micro-fibre cloth with isopropyl alcohol.
2. Hold the calibration sphere by the shaft with one hand.
3. Wipe the calibration sphere extremely carefully with the moistened cloth using only light pressure.
4. Carefully dry the calibration sphere using a dry micro fibre cloth.
Wet cleaning the calibration sphere with isopropyl alcohol
2.7.3 Inspection
A manufacturer inspection is recommended once a year. For further information, contact the service department at
m&h Inprocess Messtechnik GmbH.