2.3.4 Warming up the laser scanner
INFORMATION
The system accuracy is considered stable after 60 minutes of use (“warm-up” time).
The scanner maximum location error is 30 µm after stabilization.
The scanner actual location error is written in the "Inspection Report" delivered with the Scanner.
For reference, the subsequent table shows sample data of the difference in accuracy between the actual used
time and the stabilized accuracy, at an environment temperature of 23°C.
The system (machine with laser scanner) actual accuracy is evaluated on the machine during installation.
Time [min]
Offset Error [mm]
Depth Error [mm]
0
0,024
0,004
1
0,012
0,010
3
0,011
0,010
5
0,009
0,005
15
0,006
0,001
30
0,004
0,003
60
0,000
0,000