8.
Adjustment Status
Number of measurements
: Shows the number of runs completed.
One run consists of a measurement in face I and face II.
l Component quality (1
σ
)
: and similar lines show the standard
deviations of the determined adjustment errors. The standard devi-
ations can be calculated from the second run onwards.
☞
Measure at least two runs.
9.
Next
to continue with the check & adjust procedure.
10.
Select
Add another calibration loop
if more runs have to be
added.
Next
OR
Select
Finish the calibration & store the results
to finish the
calibration process.
Next
to view the adjustment results.
11.
Select
Finish
to accept the results. No more runs can be added
later.
OR
Select
Redo
to decline all measurements and to repeat all calibra-
tion runs.
OR
Back
returns to the previous screen.
IF the results are
THEN
to be stored
If the Use status is set to Yes,
Next
over-
writes the old adjustment errors with the
new ones.
to be determined again
Redo
rejects all new determined adjustment
errors and repeats the whole procedure.
Refer to paragraph
.
5.4
Tilting Axis Adjustment (a)
This adjustment procedure determines the following instrument error:
Instrument error
Description
a
Tilting-axis error
The following table explains the most common settings.
☞
Determine the horizontal collimation error (c) before starting this
procedure.
1.
Leica Captivate - Home
:
Settings
\
TS instrument
\
Check & adjust
2.
Check & Adjust
Select the option:
Check & adjust the tilting axis
Next step
Description
Determination of
tilting axis error step-
by-step
Check & Adjust
59