Face I measurement
If
Calibrate the automatic target aiming
is checked and an
ATRplus is available, the adjustment will include the determination of
the ATRplus Hz and V adjustment errors.
☞
Use a clean Leica standard prism as the target. Do not use
a 360° prism.
If
Calibrate the telescope camera
is checked and a telescope
camera is available, the adjustment includes the determination of
the telescope camera zero point.
☞
Use a clean Leica standard prism as the target. Do not use
a 360° prism.
5.
Aim the telescope accurately at
a target at about 100 m dis-
tance. The target must be posi-
tioned within ±9°/±10 gon of
the horizontal plane.
008701_001
± 9°
~ 100 m
6.
Measure
to measure and to
continue to the next screen.
If
Calibrate the telescope
camera
has been checked, aim
at the same target accurately
with the telescope camera using
the view finder and the digital
crosshair on the display.
Meas-
ure
to measure and to continue
to the next screen.
☞
The fine pointing has
to be performed
manually in both
faces.
008702_001
180°
180°
7.
Face II measurement
Measure
to measure the same target in the other face.
If
Calibrate the telescope camera
has been checked, aim at the
same target accurately with the telescope camera using the view
finder and the digital crosshair on the display.
Measure
measure to
the target and to calculate the instrument errors.
☞
If one or more errors are bigger than the predefined limits, the pro-
cedure must be repeated. All measurements of the current run are
rejected and none of them is averaged with the results from previ-
ous runs.
58
Check & Adjust