![Hettich ROTOFIX 46 Operating Instructions Manual Download Page 30](http://html1.mh-extra.com/html/hettich/rotofix-46/rotofix-46_operating-instructions-manual_2127810030.webp)
EN
30/87
1
Use according to specification
This device is a centrifuge which was designed only for the separation of materials or mixtures with a density of no
more than 1.2 kg/dm³ and is therefore only intended for this use. Another use or one which goes beyond this, is
considered to be non-intended. This also excludes centrifugation of materials of human origin within the framework of
clinical processes. The company Andreas Hettich GmbH & Co. KG is not liable for damage resulting from this.
Observing all information in the operating instructions and complying with the measures described therein is also a
part of the intended use.
2 Remaining
risks
The device is built according to the state-of-the-art and the recognized safety regulations. If used and handled
improperly, there could be life-threatening danger to the user or third parties, or the device could be impaired or there
could be other property damage. The device is only to be used for its intended purpose and only when it is in safe
working condition.
Malfunctions which could affect safety must be corrected immediately.
3 Technical
specifications
Manufacturer
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
Model
ROTOFIX 46
ROTOFIX 46 H
Type
4600 4600-01 4600-50 4600-51
Mains voltage (
10%)
200-240 V 1
100-127 V 1
200-240 V 1
100-127 V 1
Mains frequency
50 – 60 Hz
Connected load
460 VA
500 VA
600 VA
650 VA
Current consumption
2.5 A
5.2 A
2.5 A
5.3 A
Max. capacity
4 x 290 ml
Allowed density
1.2 kg/dm
3
Speed (RPM)
4000
Force (RCF)
3095
Kinetic energy
5700 Nm
Obligatory inspection (BGR 500)
no
Ambient conditions (EN / IEC 61010-1)
Set-up site
Indoors only
Altitude
Up to 2000 m above sea level
Ambient temperature
2°C to 35°C
Humidity
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Excess-voltage category
(IEC 60364-4-443)
Pollution degree
2
Device protection class
Not suitable for use in explosion-endangered areas.
EMC
Emitted interference,
Interference immunity
EN / IEC 61326-1,
Class B
FCC Class B
EN / IEC 61326-1,
Class B
FCC Class B
Noise level (dependent on rotor)
66 dB(A)
46 dB(A)
Dimensions
Width
538 mm
Depth
647 mm
Height
345 mm
Weight
approx. 60 kg