EN
55/133
31.3
Centrifugation run with program linking
x
Press the
PROG
key twice in order to select the parameter PR-PART. The input field will be displayed inversely.
x
Set the program location for the first program (XX+) by means of the control knob .
x
Press the
RCL
key. The centrifugation data for the selected program location will be displayed.
x
Press the
START
key. The rotation indicator will appear as long as the rotor is turning.
The run-up and braking steps for the program linking will be displayed:
x
PR-PART
First program (XX+)
x:
Run-up step for the first program
x
PR-PART
Next program (+XX+)
x:
Run-up step for the next program
y
x
PR-PART
Last program (+XX)
x:
Run-up step for the last program
y:
Braking step for the last program
x
When the time in the last program has expired, run-down will take place with the braking step of the last
program. If the centrifugation run is interrupted by pressing the
STOP
key, run-down will take place with the
braking step of the program running at the time.
31.4
Deleting program linking
x
Press the
PROG
key in order to display the parameter PROG-Nr. The input field will be displayed inversely.
x
Set the program location for the first program (XX+) by means of the control knob .
x
Press the
RCL
key. The centrifugation data for the selected program location will be displayed.
x
Press the
PROG
key twice in order to display the parameter PR-PART. The input field will be displayed inversely.
x
Press the
STO
key twice.
x
Press the
PROG
key.
32 Cooling
The temperature set-point can be adjusted from -20°C to +40°C. On centrifuges with a heating/cooling option, the
temperature set-point is settable from -20°C to +60°C. The lowest obtainable temperature is dependent on the rotor
(see Chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories"). If the actual temperature and the
set temperature deviate more than 5 °C, the temperature value in the display will start to flash on and off.
32.1 Standby-cooling
With rotor standstill and closed lid the centrifugal chamber is cooled to the pre-selected temperature. The
temperature set-point is shown in the display.
32.2
Pre-cooling the rotor
For rapid pre-cooling of the unloaded rotor and the accessories, it is recommended that the centrifuge is run with the
continuous running settings and a speed of ca. 20% of the maximum rotor speed.
33 Heating
During operation of the centrifuge the centrifuge chamber can be heated as required to the preselected temperature.
The heating is switched off when the rotor is stationary.
Risk of burning! The surface temperature of the heating element in the centrifugal chamber of the centrifuge
can reach 500°C / 932°F. Do not touch the heating element.