EN
AB1101EN_SA
Rev. 01 / 11.2022
7/45
5 Technical
specifications
Manufacturer
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
Model EBA
280
EBA
280 S
Basic-UDI-DI 4050674010008QE
Type 1101
1101-01
1102
1102-01
Mains voltage (
10%)
200 - 240 V 1
100 - 127 V 1
200
- 240 V 1
100 - 127 V 1
Mains frequency
50 - 60 Hz
50 - 60 Hz
50 - 60 Hz
50 - 60 Hz
Connected load
185 VA
185 VA
330
VA
330
VA
Current consumption
0.85 A
1.75 A
1.6
A
3.0
A
Max. capacity
6 x 50 ml
Allowed density
1.2 kg/dm
3
Speed (RPM)
6000
Force (RCF)
4146
5071
Kinetic energy
2700 Nm
Obligatory inspection
(DGUV Regel 100 - 500)
no
Ambient conditions (EN / IEC 61010-1)
Set-up site
Indoors only
Altitude
Up to 2000 m above sea level
Ambient temperature
2°C to 40°C
Humidity
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Excess-voltage category
(IEC 60364-4-443)
Pollution degree
2
Device protection class
Not suitable for use in explosion-endangered areas.
EMC
Emitted interference,
Interference immunity
EN / IEC
61326-1,
Class B
FCC Class B
EN / IEC
61326-1,
klasse B
FCC Class B
Noise level (dependent on rotor)
51 dB(A)
56 dB(A)
Dimensions
Width
326 mm
Depth
400 mm
Height
242 mm
Weight
approx. 11 kg