16 Centrifugation
In accordance with EN / IEC 61010-2-020, no persons, hazardous substances or objects may be within a
safety range 300 mm around the centrifuge during a centrifugation run.
After each centrifugation run, while the lid is unlocking, the remaining number of running cycles (centrifugation
runs) will be briefly displayed.
Example:
PROG
t/min:s
RPM
If the permissible weight difference within the rotor load is exceeded, the centrifugation run is aborted during
the start-up and the following display appears:
PROG
t/min:s
RPM
A centrifugation run can be aborted at any time by pressing the
STOP
OPEN
key.
During a centrifugation run, centrifugation parameters can be selected and changed. The changed values are
stored at program location
#
.
With the
RCF
key, you can toggle between the RPM display (RPM) and RCF display ( RCF ) at any time. If
the RCF display ( RCF ) is worked with, the centrifuging radius must be input.
The following is displayed:
PROG
t/min:s
RPM
Further operation of the centrifuge is then only possible after opening the lid.
Operation errors and malfunctions will be shown (see the chapter on "Malfunctions").
Plug in the mains switch. Switch position
.
Load the rotor and close the centrifuge lid.
16.1
Centrifugation with preset time
Example:
RPM display (RPM)
Example:
RCF display ( RCF )
RCF
Press the key to activate the RPM display
(RPM) or the RCF display ( RCF ) as
needed. The LED above the key is
illuminated if the RCF display ( RCF ) is
activated.
PROG
t/min:s
RPM
PROG
t/min:s
RPM
Input the desired centrifugation parameters (see the
"Programming" chapter).
PROG
t/min:s
RPM
PROG
t/min:s
RPM
START
PULSE
Press the key.
The centrifugation run is started.
During the centrifugation run, the rotor
speed or the resulting RCF value and
remaining time are displayed.
PROG
t/min:s
RPM
PROG
t/min:s
RPM
Continued on next page