Technical File 

File No. 

ZHTF-CE-01-015 

Rev. No. 

A.5

Non Contact Infrared Body 

Thermometer

Instructions for use

Rev. Date 

2020-07-10 

Page 

5 / 19 

 

Section 1- Overview 

Intended Use 

The  HeTaiDa  Non  Contact  Infrared  Body  Thermometers  are  designed  to  be  used  for 
intermittent measurement and monitoring of human body temperature by consumers in 
the home setting and primary care setting as screening tool. 

Description of Non Contact Infrared Body Thermometer 

  Device principle and introduction 

The HeTaiDa Non contact infrared body thermometer are hand-held, reusable, 
battery  operated  devices,  which  can  measure  human  body  temperature  on 
forehead, the skin temperature on one's forehead. 
The operation principle is based on Infrared Sensor technology. The IR sensor 
can  output  different  signal  when  measuring  different  object  temperature  or  in 
different ambient temperature, and the ASIC can turn the signal from IR Sensor 
to a digital value and display it on the LCD. 

  Description on Controls, Indicators, and Symbols 

4.

Figure 1: HTD8808E 

Infrared body thermometer

 

1. IR sensor 
2. Liquid crystal display(LCD) 
3. MODE bu on 
4. Down Arrow 

 

5. Up Arrow bu on 
6. Set 

 

7. ON/measure 

 

8. 

 Cover 

9. Surface mode 
10. Body mode 
11. Data indicator 
12. Indicator of measurement result 
13. Low Ba ery indicator 
14. Volume on/o  indicator 
15. Fahrenheit 
16. Memory Number 
17.Celsius 
18.Memory indicator 
19.Bluetooth indicator

20

20.Data transmission line

Summary of Contents for HTD8808E

Page 1: ...File No ZHTF CE 01 015 Rev No A 5 Non Contact Infrared Body Thermometer Instructions for use Rev Date 2020 07 10 Page 1 24 HeTaiDa Non Contact Infrared Body Thermometer User Manual Model HTD8808E So w...

Page 2: ...and understanding these instructions NOTICE Purchase or possession of this device does not carry any express or implied license to use with replacement parts which would alone or in combination with t...

Page 3: ...a room for 30 minutes where the temperature is between15 to 40 59 104 Remove primary batteries if equipment is not likely to be used for long time This product is not waterproof do not be immersed in...

Page 4: ...fy the device Do not use this device if you think it is damaged or notice anything unusual This device comprises sensitive components and must be treated with caution Observe the storage and operating...

Page 5: ...man body temperature on forehead the skin temperature on one s forehead The operation principle is based on Infrared Sensor technology The IR sensor can output different signal when measuring differen...

Page 6: ...ckaging Refer to operating instructions Compliance with WEEE Standard Operating Temperature DO NOT THROW AWAY Intended for multiple use Operating Humidity Indicates this device is in compliance with M...

Page 7: ...ratory Accuracy Body mode 34 0 34 9 0 3 93 2 F 94 8 F 0 5 F 35 0 42 0 0 2 95 0 F 107 6 F 0 4 F 42 1 43 0 0 3 107 8 F 109 4 F 0 5 F Surface mode 2 3 6 F Display Resolution 0 1 0 1 Three color Backlight...

Page 8: ...m o C m e t I with Equipment classification Safety Standards EN 60601 1 2006 A1 2013 EN 60601 1 2 2015 Type of protection Internally powered equipment on battery power Degree of protection Non Applie...

Page 9: ...e between the eyebrows above and keep the vertical distance press the On Scan button temperature display immediately see figure 2 When the ambient temperature of the thermometer changes too much such...

Page 10: ...e measurement read the data Note 1 After full display over you will hear a rattle or beep beep beep three times which means that the measurements have been completed while the target value of the meas...

Page 11: ...set according to the subjects of different colors and different environments data to meet the different characteristics of populations or individuals HTD8808E as below step 1 Unit Set F1 Under the boo...

Page 12: ...ng software e hardware support Bluetooth 4 0 BLE 3 Troubleshooting MESSAGE SITUATION SOLUTION Temperature taken in not within Typical human temperature range 34 0 43 0 or 93 2 109 4 Make sure the fore...

Page 13: ...g the battery be sure the system is already power off 2 Remove the battery and replace with 2 new one type AAA Make sure align properly as indicated inside the battery cover 4 Slide the battery cover...

Page 14: ...At the end of its life the appliance should not be disposed of in household rubbish Enquire about the options for environment friendly and appropriate disposal Take local regulations into account War...

Page 15: ...nformation provided and this unit can be affected by portable and mobile RF communications equipment 2 Caution Do not use a mobile phone or other devices that emit electromagnetic fields near the unit...

Page 16: ...t 8 KV Air 2 4 8 15 KV Same as the left Electrical fast transient burst IEC 61000 4 4 The input a c power ports 2 KV The input d c power ports 2 KV Signal input output ports 1 KV Not applicable Surge...

Page 17: ...pplies NOTE 2 These guidelines may not apply in all situa ons Electromagne c propaga on is a ected by absorp on and re ec on from structures objects and people a Field strengths from xed transmi ers s...

Page 18: ...mperature runs approximately from 35 5 to 37 8 95 9 F 100 F To determine if one has a fever compare the temperature detected with the person s normal temperature A rise over the reference body tempera...

Page 19: ...nsible for compliance could void the user s authority to operate the equipment This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of th...

Reviews: