Technical File 

File No. 

ZHTF-CE-01-015 

Rev. No. 

A.5

Non Contact Infrared Body 

Thermometer

Instructions for use

Rev. Date 

2020-07-10 

Page 

2 / 19 

2

Foreword 

The non contact Infrared body thermometer operating Instructions intend to provide the 
necessary information for proper operation of HTD8808E thermometer model. 
Only body mode was reviewed and certified by notified body. 
General knowledge of Infrared thermometer and an understanding of the features and 
functions of the HTD8808E thermometer model are prerequisites for proper use. 
The non contact infrared body thermometer is a medical device, and can be used 
repeatedly, which using life is 5 years. 
Please read the manual first before using it, if not fully understand the usages, please st
op using the thermometer. 

Do

not operate any of the models HTD8808E thermometer without completely 

reading and understanding these instructions.

NOTICE 

Purchase or possession of this device does not carry any express or implied license to 
use  with  replacement  parts  which  would,  alone  or  in  combination  with  this  device,  fall 
within the scope of one of the relating patents. 

For further information contact: 

Hetaida Technology Co., Ltd. 

Adds: Room 801, 802,803,804,901, 2# Building Scientific Research Center, Songhu lntelligent Valley, 
No.6 Minfu Road, Liaobu Town, Dongguan City, Guangdong Province,P.R.China

Tel: +86 0769-83326886 

Fax: +86 0769-82658050                   

0598

Contact: Tom. Chen 
E-mail: [email protected] 

About this manual edition 

Current Updated Edition/Revision: A.5, July 10, 2020;   

Safety Information 

This  device  may  only  be  used  for  the  purposes  described  in  these  instructions.  The 
manufacturer cannot be held liable for damage caused by incorrect application. 

Company Name: Share Info Consultant Service LLC Repräsentanzbüro 
Company Address: Heerdter Lohweg 83, 40549 Düsseldorf 
Tel: 0049 1767 0057022
Fax: 0049 1767 0057022
Email: [email protected]; [email protected];

Summary of Contents for HTD8808E

Page 1: ...File No ZHTF CE 01 015 Rev No A 5 Non Contact Infrared Body Thermometer Instructions for use Rev Date 2020 07 10 Page 1 24 HeTaiDa Non Contact Infrared Body Thermometer User Manual Model HTD8808E So w...

Page 2: ...and understanding these instructions NOTICE Purchase or possession of this device does not carry any express or implied license to use with replacement parts which would alone or in combination with t...

Page 3: ...a room for 30 minutes where the temperature is between15 to 40 59 104 Remove primary batteries if equipment is not likely to be used for long time This product is not waterproof do not be immersed in...

Page 4: ...fy the device Do not use this device if you think it is damaged or notice anything unusual This device comprises sensitive components and must be treated with caution Observe the storage and operating...

Page 5: ...man body temperature on forehead the skin temperature on one s forehead The operation principle is based on Infrared Sensor technology The IR sensor can output different signal when measuring differen...

Page 6: ...ckaging Refer to operating instructions Compliance with WEEE Standard Operating Temperature DO NOT THROW AWAY Intended for multiple use Operating Humidity Indicates this device is in compliance with M...

Page 7: ...ratory Accuracy Body mode 34 0 34 9 0 3 93 2 F 94 8 F 0 5 F 35 0 42 0 0 2 95 0 F 107 6 F 0 4 F 42 1 43 0 0 3 107 8 F 109 4 F 0 5 F Surface mode 2 3 6 F Display Resolution 0 1 0 1 Three color Backlight...

Page 8: ...m o C m e t I with Equipment classification Safety Standards EN 60601 1 2006 A1 2013 EN 60601 1 2 2015 Type of protection Internally powered equipment on battery power Degree of protection Non Applie...

Page 9: ...e between the eyebrows above and keep the vertical distance press the On Scan button temperature display immediately see figure 2 When the ambient temperature of the thermometer changes too much such...

Page 10: ...e measurement read the data Note 1 After full display over you will hear a rattle or beep beep beep three times which means that the measurements have been completed while the target value of the meas...

Page 11: ...set according to the subjects of different colors and different environments data to meet the different characteristics of populations or individuals HTD8808E as below step 1 Unit Set F1 Under the boo...

Page 12: ...ng software e hardware support Bluetooth 4 0 BLE 3 Troubleshooting MESSAGE SITUATION SOLUTION Temperature taken in not within Typical human temperature range 34 0 43 0 or 93 2 109 4 Make sure the fore...

Page 13: ...g the battery be sure the system is already power off 2 Remove the battery and replace with 2 new one type AAA Make sure align properly as indicated inside the battery cover 4 Slide the battery cover...

Page 14: ...At the end of its life the appliance should not be disposed of in household rubbish Enquire about the options for environment friendly and appropriate disposal Take local regulations into account War...

Page 15: ...nformation provided and this unit can be affected by portable and mobile RF communications equipment 2 Caution Do not use a mobile phone or other devices that emit electromagnetic fields near the unit...

Page 16: ...t 8 KV Air 2 4 8 15 KV Same as the left Electrical fast transient burst IEC 61000 4 4 The input a c power ports 2 KV The input d c power ports 2 KV Signal input output ports 1 KV Not applicable Surge...

Page 17: ...pplies NOTE 2 These guidelines may not apply in all situa ons Electromagne c propaga on is a ected by absorp on and re ec on from structures objects and people a Field strengths from xed transmi ers s...

Page 18: ...mperature runs approximately from 35 5 to 37 8 95 9 F 100 F To determine if one has a fever compare the temperature detected with the person s normal temperature A rise over the reference body tempera...

Page 19: ...nsible for compliance could void the user s authority to operate the equipment This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of th...